Coincidence-Summing corrections in high resolution gamma-ray spectrometry: simplified analytical expressions
- Autori: Tomarchio, EA
- Anno di pubblicazione: 2010
- Tipologia: Proceedings
- Parole Chiave: Coincidence-summing; gamma-ray spectrometry
- OA Link: http://hdl.handle.net/10447/57343
Abstract
Simplified expressions to determine the coincidence-summing correction factors by using a matrix representation of a decay scheme are reported. Expressions can be used directly for point sources, with the introduction of full-energy-peak and total efficiencies, while for extended sources a more complex procedure, though approximate, is suggested.