Image-based MRI Gradient Estimation
- Authors: Acquaviva, R.; Mangione, S.; Garbo, G.
- Publication year: 2018
- Type: Articolo in rivista (Articolo in rivista)
- OA Link: http://hdl.handle.net/10447/254010
Abstract
In order to reduce geometric distortion phenomena in MR images, every MRI system main magnet undergoes a shimming process. Since this process aims at optimizing magnetic field homogeneity within a so-called uniformity sphere, image quality outside this sphere is neglected. Since the fields vary smoothly in space, MR signal-to-noise ratio is still non-zero just outside the uniformity region, but correction of MR image distortion fails due to lack of magnetic field knowledge outside it. We propose a novel algorithm for measuring all the fields involved in the generation of images. Our proposal is based on exploitation of the distortion which can be observed in images of a known phantom. The proposed method will enable measurement of the fields in a region that can be bigger than the uniformity sphere depending on the phantom dimensions.