Salta al contenuto principale
Passa alla visualizzazione normale.

CATERINA MARIA GAMBINO

Comparative analysis of biochip mosaic-based indirect immunofluorescence with enzyme-linked immunosorbent assay for diagnosing myasthenia gravis

  • Autori: Gambino C.M.; Agnello L.; Lo Sasso B.; Scazzone C.; Giglio R.V.; Candore G.; Ciaccio A.M.; Di Stefano V.; Brighina F.; Vidali M.; Ciaccio M.
  • Anno di pubblicazione: 2021
  • Tipologia: Articolo in rivista
  • OA Link: http://hdl.handle.net/10447/527321

Abstract

Background: The detection of anti-acetylcholine receptor (AChR) and anti-muscle-specific tyrosine kinase (MuSK) antibodies is useful in myasthenia gravis (MG) diagnosis and management. BIOCHIP mosaic-based indirect immunofluorescence is a novel analytical method, which employs the simultaneous detection of anti-AChR and anti-MuSK antibodies in a single miniature incubation field. In this study, we compare, for the first time, the BIOCHIP MG mosaic with conventional enzyme-linked immunosorbent assay (ELISA) in the diagnosis of MG. Methods: A total of 71 patients with MG diagnosis were included in the study. Anti-AChR and anti-MuSK antibodies were measured separately by two different ELISA and simultaneously by BIOCHIP. The results were then compared. Results: The overall concordance between ELISA and BIOCHIP for anti-AChR reactivity was 74%. Cohen’s kappa was 0.51 (95% CI 0.32–0.71), which corresponds to 90% of the maximum possible kappa (0.57), given the observed marginal frequencies. The overall concordance for anti-MuSK reactivity was 84%. Cohen’s kappa was 0.11 (95% CI 0.00–0.36), which corresponds to 41% of the maximum possible kappa (0.27). Conclusion: The overall concordance among assays is not optimal.